Firmware compatibility table
Please, alway make sure a firmware is compatible with your hardware before upgrading.
Firmware version
Hardware compatibility
-
PRO One version 2
Release Notes
Firmware GUI version 1.30
New Features:
- Cross Mode feature added: The Cross Mode allows the use of four electrodes connected to the two outputs.
Output 1 delivers frequency A, while output 2 delivers frequency B.
Cross Mode is available in both Manual Mode and Protocol Mode.
- Continuous intensity adjustment: A long press on the intensity “+” and “-” buttons now continuously adjusts the intensity.
- Improved navigation between the screens: The “<” button now enables the user to return to the previous screen.
- Field editing enhancement: When editing a field, if the field is empty and the “Enter” button is pressed, the previous value is restored.
- Automatic channel selection with frequency editing: Pressing frequency buttons A, B, C, or D now automatically positions the channel selector.
- Sleep mode update: A fade effect has been added when the screen enters sleep mode, differentiating it from screen shutdown.
- Screen flicker reduction: The screen flicker has been further minimized, providing an even smoother visual experience.
Bug Fixes:
#B0008: A protocol may pause unexpectedly.
#B0009: On the protocol selection screen, the displayed times may not always be accurate.
#B0010: Opening an empty folder caused the PRO One to crash.
#B0011: After opening a protocol and returning to Manual Mode, the Polarity and Waveform fields were not reset to their default values.
#B0012: The active polarity was not updated when navigating within a protocol.
#B0013: The active waveform was not updated when navigating within a protocol.
#B0014: In Perfect Editor software v1.3, adding a space before or after the frequencies caused the PRO One to interpret the empty frequency as 0. This issue has been resolved in the PRO One and in the Perfect Editor software version 1.4.